Shortening
The cited HBK PW/SP manual distinguishes its four-wire cable calibration from six-wire configurations. Confirm your own supplied model's instructions and recalibration requirements before cutting a cable.
ENGINEERING WORKSHEET
Identify a chart and record your installation for support.
Go to the inputs and results ↓WIRING + OBSERVATIONS
Before mechanical inspection or resistance testing, unload and isolate the equipment using the approved site procedure. Qualified personnel should use the actual sensor and instrument manuals. Never change wiring while energized.
Unknown model: no wire colours or jumpers are assigned. Send the full code, nameplate, source cable chart and indicator terminal document.
Select an observed behavior to see the evidence checklist. This tool does not classify a sensor as damaged.
Carry these observations into the support RFQFUNCTION BEFORE COLOUR
| Configuration | Functions normally exposed | Review before connection |
|---|---|---|
| Analog four-wire | EXC+, EXC−, SIG+, SIG−; shield separately | Cable calibration, exact functions, instrument input type and any required manual-specific links |
| Analog six-wire | Four bridge functions plus SEN+ and SEN−; shield separately | Whether the instrument supports sense, actual terminations and the manufacturer-approved connection arrangement |
| Connector/digital/unknown | No analog pinout inferred from conductor count | Actual protocol, supply and connector pin assignment; leave the chart unassigned until identified |
These are identified source examples. They demonstrate why a generic manufacturer-name colour lookup cannot establish an installed connection.
| Specific reference | Blue conductor in this source | Source scope |
|---|---|---|
| SPL-SSW-101 | EXC+ | Own reference chart, SPL-REF-20261122.1 |
| SPL-HVY-101 | SEN− | Own reference chart, SPL-REF-20261122.1 |
| Original Flintec PC6 six-conductor option | SEN+ | A28-Rev11-GB, PDF page3; verify the supplied option |
Do not merge these charts into a colour-to-colour replacement instruction. Compare the original and proposed function labels, polarity, cable/connector option, excitation and instrument requirements independently.
Review original-model electrical and mechanical interfaces
Confirm the fixed/loading ends, clean mounting faces, spacer contact and permitted fastener engagement. Keep the active beam, cable, platen and stops clear through the approved working travel. Contact or debris can create an unintended load path.
Review plate flatness and rigidity against the actual source drawing. This guide does not assign a universal flatness value or fastener grade. Preserve the drawing revision and assembly inspection record.
Review spacer and overload-base principles| Original source example | Conditions stated by that source | Boundary for an SPL installation |
|---|---|---|
| HBK PW15AH,10–100 kg | M6; minimum property class10.9;14 N·m reference, with screw-manufacturer dimensions considered | A02435 03 Y00 02, English section4.2 / PDF page10; original model only |
| Flintec PC6 | M6 class8.8;10 N·m; source explicitly assumes oiled threads | A28-Rev11-GB, PDF page3; original source conditions only |
| Current SPL references | Thread/depth data available; approved joint torque not supplied | Request the exact bolt, lubrication, engagement, plate and tightening procedure; neither original example is assigned to SPL |
The cited HBK PW/SP manual distinguishes its four-wire cable calibration from six-wire configurations. Confirm your own supplied model's instructions and recalibration requirements before cutting a cable.
Record cable type/length, splice position, shield continuity, conductor functions and environmental protection. Sense support and connection location depend on the instrument and sensor arrangement.
Review routing near drives/power wiring and the local shield/earth scheme. A shield is not a substitute for an engineered protective-earth or potential-equalization path.
HBK PW/SP A02323, English section5 provides the identified original configuration examples; follow the actual supplied manuals for implementation.
| SPL reference | Input / output resistance | Cable reference | Source chart |
|---|---|---|---|
| SPL-CMP-101 | 405 ±5 Ω / 350 ±3 Ω | 4 mm diameter × 350 mm; 4 wires plus shield | Open reference PDF |
| SPL-CMP-102 | 380 ±40 Ω / 380 ±40 Ω | 4 mm diameter × 400 mm; 4 wires plus shield | Open reference PDF |
| SPL-SSW-101 | 420 ±20 Ω / 350 ±10 Ω | 5.2 mm diameter × 1,600 mm; 6 wires plus shield | Open reference PDF |
| SPL-ALU-101 | 420 ±20 Ω / 350 ±10 Ω | 5.3 mm diameter × 2,000 mm; 6 wires plus shield | Open reference PDF |
| SPL-HVY-101 | 380 ±20 Ω / 350 ±10 Ω | 6.3 mm diameter × 2,500 mm; 6 wires plus shield | Open reference PDF |
| SPL-SSW-102 | 380 ±20 Ω / 350 ±10 Ω | 6 mm diameter × 3,000 mm; 6 wires plus shield | Open reference PDF |
| SPL-SSW-103 | 390 ±5 Ω / 350 ±5 Ω | 6 mm diameter × 1500 mm; 6 conductors plus shield | Open reference PDF |
Qualified personnel should isolate the transducer from the instrument and other electrical paths according to both manuals before resistance testing. Compare the correct EXC or SIG pair and the actual model specification. An in-circuit reading or one unexpected value does not independently establish a failed sensor.
Insulation testing is a separate procedure. No voltage, threshold or connection arrangement is preselected here; obtain the approved model/electronics procedure before testing.
| Observed behavior | Possible causes | Evidence to retain |
|---|---|---|
| Position-dependent / corner error | Unintended contact, frame distortion, mounting interface, eccentric load outside the approved envelope or calibration conditions. | Record centre and corner readings using the approved load/placement plan, plus pre/post zero. Photograph spacers, stops and cable clearance. Do not file the sensing beam. |
| Zero drift / unstable reading | Temperature change, settling product, air movement, cable strain, unintended contact, moisture or electrical interference. | Record time, ambient/process temperature, load state and whether nearby equipment was operating. Inspect clearances and visible cable damage with the equipment isolated. |
| No reading / instrument input error | Instrument setup, cable/connector interruption, mismatched functions, excitation compatibility or an actual bridge fault. | Capture the instrument model, error code, approved terminal chart, full sensor code and whether the cable is four- or six-wire. Do not identify terminals by colour alone. |
| Reading direction is reversed | Applied load direction, instrument polarity/configuration or an incorrect signal-function connection. | Record the applied direction and known-load response, sensor orientation and the actual terminal-function chart. |
| Change after cleaning / moisture exposure | Cable entry/connector exposure, retained liquid, chemical/temperature effects, contamination contact or sensor damage. | Record cleaning temperature, pressure, chemistry, timing and installation drainage; photograph the entry and connector without opening sealed sensor parts. |
| Change after impact / overload | Shifted stops/mounting, cable damage, mechanical obstruction or permanent sensor/frame damage. | Unload and isolate the equipment. Record the event, previous calibration/zero and visible assembly condition. Preserve the full model code and installation photos. |
The Flintec LCT-11 guide distinguishes cable, bridge, moisture and other possible causes in its own instrument workflow. Its tester limits and multi-cell examples do not establish acceptance limits for an isolated single-point installation.
Agree the model-specific permitted mass, positions, settling conditions and error limits before testing. The following is a recording framework, not a universal statutory test or permission to load beyond the rated envelope.
| Record position / state | Record with each reading | Decision boundary |
|---|---|---|
| Unloaded reference | Stable zero, time, temperature, instrument/filter settings | Identify warm-up, settling and assembly conditions |
| Centre + four agreed off-centre positions | Applied mass, placement coordinates, settled reading and repeats | Use the actual model/assembly's permitted loading and agreed acceptance criteria |
| Unload after the sequence | Return-to-zero, elapsed time and visible contact/cable changes | Investigate differences with source-specific controlled checks |
Keep the full order code, capacity, cable option and approval markings readable. Type the code in the inquiry so every suffix survives the handoff.
Show fixed/load pads, spacers, stops, supports and cable exit. Add the dimensioned assembly drawing; a photo cannot establish hole pitch, gap or thread depth.
Capture the instrument model, terminal-function labels and approved wiring chart. Include error/event history and previous calibration records where available.
Attach approved files in your email application. Ask about an NDA before sending confidential drawings. A damaged-cell replacement should follow identified source/code and interface review, rather than a promised universal swap or delivery time.
Open the original-model review centreSOURCE REVIEW / 22 NOVEMBER 2026
The SPL reference PDFs use the reviewed source catalogue revision. Independent manufacturer documents below support the identified original examples and installation distinctions; they do not assign their ratings, torque, pinout or approvals to an SPL configuration.
No. Confirm the exact supplied model, cable/connector option and instrument terminal-function chart. The same colour can mean excitation or a sense function in different references; a brand name or cable count alone is not a connection instruction.
A typical analog four-wire bridge exposes excitation plus/minus and signal plus/minus; a six-wire configuration adds sense functions for the instrument. The shield is separate from those counts. Actual terminations, sense support and any bridging must follow both supplied manuals.
No. Identify the sensor functions and instrument architecture first. Some documented original configurations specify same-polarity sense/excitation connections or amplifier-side links, but their location and conditions are manual-specific. The chart tool does not prescribe a universal jumper.
Check the actual model and calibration conditions. The cited HBK PW/SP instructions distinguish four-wire cable calibration from six-wire sense configurations. Extension type, termination, shielding, excitation and recalibration requirements still need review; never infer that every six-wire cable modification is unrestricted.
No. The original PW15AH and PC6 examples have different fastener/torque conditions despite M6 callouts. Use the exact supplied sensor and joint specification, including bolt grade, lubrication, usable engagement and plate material. No universal SPL torque is assigned here.
Use the approved test mass and placement plan within the actual model/assembly limits. Record the source drawing, capacity, platen dimensions, centre and four position readings, pre/post zero, instrument settings, environment and agreed error criteria. This is a record framework, not a universal legal-metrology test method.
No. Mechanical contact, product movement, temperature, cable strain, interference, instrument configuration and damage can produce overlapping symptoms. Record the event and conditions, then follow the model-specific isolation and testing procedure before deciding.
No. Input and output resistances are separate specifications and depend on the design. Use the selected reference PDF and actual order revision; instrument connections and parallel paths can alter a measurement. An out-of-range observation needs a controlled isolated test and source comparison.
No. Agree the manufacturer-approved procedure, voltage, connections and instrument isolation before any insulation test. Do not apply a generic high-voltage test through connected electronics or infer an SPL test condition from another manufacturer sheet.
Leave the SPL chart unselected and enter the complete installed code. Send the nameplate, mounting and indicator-terminal photos with the source diagram. The support request retains your observations without assigning a wire chart or confirmed replacement.
DRAWING + INPUTS + ACCEPTANCE SCOPE
Use the tool's review action to retain your entered values in the editable engineering brief. Attach the approved drawing, nameplate or test records in your own email application.
Return to my worksheet / observationsStart a blank installation inquiry
The calculations and symptom lists organize a review. Agree the actual supplied model, assembly, source revisions and tests before release.